Baby Infant Radiant warmer Brief Introduction:.
Two control modes:Air mode and baby mode controlled by micro-computer.
Humidity concentration servo control function and oxygen concentration display function.
Humidity concentration and oxygen concentration servo control function (for GT550-103)
Set temperature, air temperature, baby temperature, humidity concentration are displayed separately.
>37'C temperature set function.
Failure alarm indication.
Embedded integrated sensor box.
Integrated water reservoir in the shape of drawer.
Double wall acrylic glass hood, automatic air circulation device.
Inclination of bassinet is adjustable.
RS-232 connector..
Baby Infant Radiant warmer Standard configuration:
BB-100N:
Main body (including the acrylic glad hood, temperature controller,
bassinet), skin temperature sensor, I.V. pole, mattress, guard rail,
mobile cabinet.
BT-100: Based on BB-100N, with humidity servo control.
Baby Infant Radiant warmer Specification:.
Power supply: AC220V-230V,50HZ
Power input: 1000VA
Mode of control: Air mode and baby mode controlled by micro-computer
Control range of air temperature: 25°C-37°C (override mode 37-39°C)
Control range of baby temperature: 34°C-37°C (override mode 37-38°C)
Accuracy of skin temperature sensor: ≤0.3°C
Variability of temperature: ≤0.5°C
Uniformity of temperature: ≤0.8°C(mattress is on the horizontal position)
Inclination of the bassinet: ±12°
Noise indise hood: ≤60dB(A)
Inclination of bassinet: 0°~5°
Faolure
alarm: over temperature alarm, deviation alarm, sensor failure alarm,
fan motor alarm, power failure alarm, sensor box position alarm, low
water alarm, tank position error alarm and system alarm.
Indication range of humidity: 0%RH~99%RH
Control range of humidity :0%RH~90%RH
Accuracy of humidity: ±10%RH
Indication range oxygen concentration: 0%-99%
Control range of oxygen concentration:20%~60%(for GT550-103)
Accuracy of oxygen concentration:±3%.