Model: HIP-EA3
01. DPS(动态压力系统)确保实时闭塞压力。
02.特别原创双时尚,密集和正常的营养模式
03.在工作期间不要更改预设数据,以免误操作,提高进料安全性
04. USB用于更新软件和记录历史数据
| Model No. Parameters |
Enteral Feeding Pump |
| Flow rate range(ml/h) | 1~400, step by 1ml/h |
| Accuracy of flow rate | ±10%, with dedicated IV set |
| volume limit(ml) | 0-9999, step by 1 ml |
| Accuracy of infusion volume | ±10%, with dedicated IV set |
| Prime rate and accuracy(ml/h) | 1000 ±10% |
| Flush rate and accuracy(ml/h) | 1000 ±10% |
| One-time flush volume(ml) | 40±10% |
| Max infusion pressure | ≥0.1 MPa |
| Occlusion alarm value(pressure) | 50KPa±10KPa |
| IV set | Refer to the recommended IV set list |
| Bubble sensitivity | Detestability of air bubble≥50uL |
| Alarming | Occlusion, Battery exhausted |
| Indicating functions | Bubble in line, VTBI completion, Flush completion, Low battery, Malfunction(ER series),Pause overtime, Power interruption, Infusion mode, Working state |
| Noise | With equipment operating on a wooden table in the maximum flow rate, and in the distance of 1 meter in every direction, the noise level is below 50dB. |
| Power | AC:100V-240V 50 Hz /60Hz |
| Internal power | After fully charged, internal DC11.1V rechargeable Li battery can run over 5 hours at intermediate rate |
| Weight | 2.5kg |
| Classification | Type BF, Class II continuous volumetric feeding pump,IPX1(sealed drip-proof device) |
| Size | 170mm(W)×155mm(L)×180mm(H) |
| Optional Accessory | Fluid warmer, Dedicated disposable nutrition IV set |